LCMS - Metanephrines

Metanephrines Reagents Set

SKU: 1060 KIT M MET

The Diagnotix kit for the determination of Metanephrines in (EDTA) plasma.

The Diagnotix Metanephrines Reagent Set is able to accurately quantify the 3-MT group.

Components

  • Metanephrine
  • Normetanephrine
  • 3-methoxytyramine (3-MT)

At a glance

  • Matrix: (EDTA) plasma
  • Technique: LC-MS/MS
  • Internal Standard: Metanephrine-D3, Normetanephrine-D3, 3-MT-D4

Included in kit (200 assays):

  • 1061 CAL M MET, Calibrator Set (Calibrator 1 – 6), 6 x 2 x 2 ml
  • 1069 M MET, Internal Standard D3 D4, 4 x 6 ml
  • 1070 M MET, Mobile Phase I, 1 x 250 ml
  • 1071 M MET, Mobile Phase II, 2 x 500 ml
  • 1072 M MET, Washing Solution 1, 2 x 220 ml
  • 1073 M MET, Washing Solution 2, 2 x 220 ml
  • 1074 M MET, Washing Solution 3, 2 x 100 ml
  • 1078 M MET, Elution Solution, 2 x 30 ml
  • 1079 M MET, Buffer, 4 x 30 ml
  • 10601 M MET, Autosampler Washing Solution, 1 x 1000 ml

The Free Metanephrines test in (EDTA) plasma measures three metabolites of the catecholamines epinephrine, norepinephrine and dopamine. These catecholamines are made in the adrenal gland to regulate the heartbeat, blood pressure and glucose concentrations. They can also be released in excess by rare tumors, such as pheochromocytomas and paragangliomas, on adrenal and extra-adrenal chromaffin tissue. The majority of these tumors release epinephrine and norepinephrine. Some rare tumors however excrete mostly, or sometimes solely, dopamine.

Chromatograms

Technical data

Applicable for at Least: Waters TQ-Sµ, Shimadzu LC-8050, Sciex Citrine
Limit of Quantification; Metanephrine / Normetanephrine / 3-MT (nmol/l) 0.032 / 0.041 / 0.093
Linearity; Metanephrine / Normetanephrine / 3-MT (µmol/l) 18 / 25 / 25
Accuracy Passes (<15% CV)
Simple Precision; Metanephrine / Normetanephrine / 3-MT (CV %) 1.6 / 1.9 / 2.8
Complex Precision; Metanephrine / Normetanephrine / 3-MT (CV %) 6.1 / 5.3 / 5.6
Ionization Electrospray
MS/MS mode Positive
Sample Preparation SPE
Preferred Column Column Acquity UPLC BEH Amide 1.7µm 2.1 x 100mm
Analysis Time 4 minutes
Flow Rate 0.4 l/min
Injection Volume 5-20 µl
Validation Report Available Yes

Please Note:

Information as presented freely on this website is not sufficient for proper operation of these medical devices. Please refer to the applicable instructions for use and all accompanying information such as warning notices. All necessary documentation can be found in the download centre section of this website.

For other LC-MS/MS systems than listed above, please contact us to discuss the possibilities.

Interested in our products or services?
Contact us
or call us on +31 (0)596 20 10 62

More info?

* Mandatory fields

Contact us for more information about this product using the contact form or the contact information below.

Contact details

De Plassen 4
9902 SE Appingedam
The Netherlands

info@diagnotix.com +31 (0)596 20 10 62