Customized solutions, better diagnostics

Vitamin A&E LC-MS

Vitamin A & E LC-MS Reagens Kit

SKU: 1080 KIT M VAE

Advantages of the Diagnotix Vitamin A&E LC-MS kit:

  • 40 times deproteinization solution instead of the usual 2-3 times, leading to increased column stability and lifespan.
  • Six Calibrator Levels in Human Serum
  • Easy to install onto your LC-MS system
  • No concern of coeluting other Tocopherols or Tocotrienols (See below the text 'Why transition from HPLC to LC-MS?' to understand why this is an important topic)

Components:

  • Vitamin A (Retinol)
  • Vitamin E (α-Tocopherol)

At a glance:

  • Matrix: serum/plasma
  • Technique: LC-MS/MS
  • Internal Standard: Retinol-D6, α-Tocopherol-D3

Intended use



The Diagnotix Vitamin A&E LC-MS kit is developed to determine the patient's Retinol and/or α-Tocopherol concentration in serum or plasma as a basis for further diagnostics.


Technical benefits


  1. only 25 µl sample needed
  2. No concern of coeluting any other vitamin E within our HPLC-assay. Commercial supplements, containing all vitamin E molecules, are widely available. This highlights the importance of using a robust HPLC-UV or fluorescence method for accurate analysis.
  3. No need for additional HPLC systems in your lab. Easily integrate this method into your existing LC-MS platform.


Why moving over from HPLC to LC-MS?


Save time and reduce maintenance costs by eliminating your HPLC system.

While analyzing vitamins A and E by HPLC (UV or fluorescence) is relatively straightforward, it requires maintaining an HPLC system. If your LC-MS system is not fully utilized, transferring your vitamin A and E method to LC-MS could save you time and reduce maintenance costs.

No need to separate other tocopherols and tocotrienols, which could potentially lead to longer runtimes and reduced column lifespan.

Another compelling reason to switch to LC-MS is the growing availability of supplements containing a wide range of tocopherols and tocotrienols analogs in varying concentrations. These can elevate the levels of other tocopherols and tocotrienols in the patient's sample, all of which share similar hydrophobic properties. While HPLC can separate these compounds, it potentially extending sample run times and becomes increasingly challenging as the column ages. Thus your column life time might be drastically reduced.
In contrast, LC-MS eliminates the need for full chromatographic separation since the molecules differ in mass. This ensures no interference from other tocopherols or tocotrienols, while maintaining a fast and efficient sample-to-sample throughput.

An enhanced deproteinization step for greater stability and increased lifespan of your column.

As a final point, with the Diagnotix LC-MS solution, you perform a significant additional dilution (40 times) during the deproteinization step, removing even more protein compared to traditional HPLC methods (which typically use 2 to 3 times dilution). This further contributes to longer column stability and lifespan.


Especially interesting for laboratories that:


I. Runs large numbers of samples once or more times per month.
II. Or are looking to transition from HPLC to LC-MS to enhance operational efficiency and reduce overall maintenance costs.

Contact Us for More Information


Interested in learning more about our Vitamin A&E LC-MS solution? Contact us via the contact form below. We looking forward to hear from you!

Contact form Webinar Modernizing LC-MS workflows: Case study overcoming Vitamin A&E challenges and embracing innovation

Included in kit (300 assays):

  • 2001 CAL HM VAE, Calibrator Set (Calibrator 1 – 6),
  • 1090 M VAE, Deproteinization Solution with Internal Standard, 3 x 100 ml
  • 1091 M VAE, Mobile Phase I, 1 x 250 ml
  • 1092 M VAE, Mobile Phase II, 1 x 500 ml
  • 1096 M VAE, Autosampler Washing Solution, 1 x 1000 ml

Chromatograms

Technical data

Applicable for at Least: Waters TQ-Sµ, Shimadzu LC-8050, Sciex Citrine
Limit of Quantification; Retinol / α-Tocopherol (µmol/l) 0.052 / < 0.927
Linearity; Retinol / α-Tocopherol (µmol/l) > 9 / 145
Simple Precision; Retinol / α-Tocopherol (CV %) 1.6 / 1.6
Complex Precision; Retinol / α-Tocopherol (CV %) 2.7 / 3.4
Ionization Electrospray
MS/MS mode Positive

Important Notice:


The information provided on this website alone is not sufficient for the correct and safe operation of these medical devices. Always refer to the applicable Instructions for Use and accompanying documentation, including any warnings and safety notices. All essential documentation is available in the Download Centre on this website.


Please note that system-specific conditions may influence results. If you have any questions or require technical support, we encourage you to contact us to discuss your specific setup or requirements.

Interested in our products or services?
Contact us
or call us on +31 (0)596 20 10 62

More info?

* Mandatory fields

Contact us for more information about this product using the contact form or the contact information below.

Contact details

De Plassen 4
9902 SE Appingedam
The Netherlands

info@diagnotix.com +31 (0)596 20 10 62

Related products